CYGNI

Power Device Parasitics Analyser

Equipment

Cosmic’s Cygni measures the full set of small-signal parasitics of a power semiconductor – Ciss, Coss, Crss, Cgs and gate resistance – from 100kHz to 5MHz with programmable drain bias to 3kV in a single device insertion. It uses the same measurement core as Cosmic’s M2 and VIP Ultra production platforms, to seamlessly move between characterisation testing and production screening.

Reasons Why

Complete parasitic set

Ciss, Coss, Crss, Cgs and Rg in one insertion, without re-cabling and under one calibration.

Relevant frequency range

Programmable from 100kHz to 5MHz, including the primary 1MHz datasheet correlation point.

Bias and precision

Drain bias to 3kV, gate bias to ±30V, bridge-null Rg measurement and software-selectable 4-terminal Kelvin sensing.

Production-ready speed

Under 30ms in high-speed mode, with measurement time reported alongside every reading.

Bench-to-line correlation

The same measurement core as Cosmic's M2 and VIP Ultra, with LXI-compatible LAN, SCPI and an embedded web interface.

Product Presentation

Power-semiconductor datasheets specify a set of small-signal parasitics – input, output and reverse-transfer capacitance, gate-source capacitance, and gate resistance. Verifying them on a general-purpose LCR meter means four-terminal configurations, four re-cablings and four separate fixture corrections, and the results will still not match the production tester used to screen the parts in volume.

Cygni closes that gap. It is a digital auto-balancing bridge requiring only one calibration cycle that measures the whole parasitic set in a single device insertion, from 100kHz to 5MHz, with a programmable drain bias to 3kV to allow capacitance to be characterised over the DUT’s full voltage range. Because the reading is taken at the bridge null, the phase resolution that gate-resistance extraction depends on is physically available – a matter of measurement architecture, yielding more sensitive measurements.

For production installations, measurements run in under 30ms in high-speed mode and the instrument is built for a minimal handler footprint with a height of just 60mm and the option of soft-dock or hard-dock DUT connection.

Control is via Cosmic’s Linux-based System Software which automates the whole production test cell, whether that be wafer sorting, KGD or packaged device testing. For characterisation use, the interface is LXI-compatible LAN with SCPI commands, or through a comprehensive stand-alone embedded web interface that needs no software installed.

FEATURES

CYGNI

Number of test sites
1 site
Consumption
≤60W (typical 30W) Input power: 90–264VAC, 47–63Hz
Parallelism
Option for two devices tested sequentially via internal switch matrix
Temperature Control
None (no DUT temperature control)
DUT connections
Single device insertion
Stand-alone operation
Browser-based, no software installation required
Compensation and hand-test fixture
Residual fixture inductance and resistance characterised and corrected via calibration
Interfaces
Embedded web interface
SCPI over LAN
LXI-compatible LAN
Production software: Cosmic HPT Linux System Software

Impedance Measurement

Measurement principle
Digital auto-balancing bridge
Measured parameters
Ciss, Coss, Crss, Cgs and gate resistance (Rg)
Frequency range
100kHz to 5MHz
Terminal configuration
Software-selectable 4-terminal Kelvin sensing
Capacitance range
3pF – 100nF
Ciss / Coss accuracy
±(2% + 2pF) absolute accuracy @1MHz, factory calibrated (provisional)
Crss accuracy
As Ciss / Coss
Gate resistance accuracy
Measured for Cg ≥1nF ±(1% + 0.05Ω) in high-repeatability mode ±(2% + 0.3Ω) absolute accuracy @1MHz, factory calibrated (provisional)
Measurement time
Under 30ms per measurement (high-speed)
Measurement time reporting
Measurement time reported with each reading
Calibration
Factory calibrated
Verification
Traceable verification using multiple reference values and frequencies
Diagnostics
Built-in self-test Internal thermal and supply monitoring
Automatic protective shutdown

Analog/Power

Drain High Voltage Power Supply
0–3kV programmable Forcing accuracy: ±(0.5V + 0.1% of setting)
Gate Voltage Power Supply
±30V programmable Forcing accuracy: ±3mV

DC Measure

Accuracy
Drain readback: typical ±1% Gate readback: typical ±5mV
Resolution
Drain: 100mV
Gate: 1.2mV

Measurements instruments

Digitizer
Coherent digital acquisition of both bridge nodes, with live waveform display
LCR meters
Digital auto-balancing bridge, 100kHz to 5MHz

Dimensions

D⨯W⨯H
330 mm ⨯ 407 mm ⨯ 60 mm
Weight
6.5kg

System

Certifications
SEMI S2/S8 compliance (pending)
CE / UKCA (pending)

More info